| MOQ: | 1 set |
| Price: | Negotiable |
| Standard Packaging: | Plywood packing |
| Delivery Period: | 5~8 working days |
| Payment Method: | L/C,T/T |
| Supply Capacity: | 500 sets/month |
| Feature | AAS-361MC | AAS-361CRT |
|---|---|---|
| Type | Flame type (air-acetylene), intelligent | Flame type (air-acetylene), intelligent |
| Optical System | Single beam, C-T type monochromator, 1200 lines/mm grating, 250nm blaze wavelength, 0~2nm continuous spectral bandwidth, F/5.4 relative aperture | Single beam, C-T type monochromator, 1200 lines/mm grating, 250nm blaze wavelength, 0~2nm continuous spectral bandwidth, F/5.4 relative aperture |
| Light Source System | Two sets of lamp power supplies, independent power supply, pulse lighting, 0~30mA lamp power | Two sets of lamp power supplies, independent power supply, pulse lighting, 0~30mA lamp power |
| Microcomputer Specifications | 8031CPU microcontroller, EPROM16KB RAM 8KB | General personal computer (premium brand, multimedia configuration) |
| Reading Display | Four-digit LED digital display, direct reading T.ACRSD; print data and reports, draw working curve recorder (2V, 1V) | Screen display reading: five-digit floating point, direct reading TACRSD; display working curve analysis report; instantaneous recording curve (paperless recorder function) |
| Working Modes | Transmittance, atomic absorption method, flame emission method | Transmittance, atomic absorption method, flame emission method |
| Signal Forms | Continuous reading (instantaneous value), average reading (hold), peak height, peak area | Continuous reading (instantaneous value), average reading (hold), peak height, peak area |
| Concentration Calculation | Linear regression, curve fitting, standard addition method | Linear regression, curve fitting, standard addition method |
| Standard Samples | 1~6 pieces | 1~8 pieces |
| Reading Preparation Time | 0~100 seconds, 0.1 second intervals | 0~100 seconds, 0.1 second intervals |
| Integral Sampling Time | 0~100 seconds, 0.1 second intervals | 0~100 seconds, 0.1 second intervals |
| Scale Extension | 1~20 times interval | 1~20 times interval |
| Average Repetitions | 1~100 times to get average value | 1~20 times |
| Result Storage | None | Available |
| Additional Functions | Automatic blanking, sensitivity and baseline drift correction, relative standard deviation calculation, automatic printing of analysis data, working curves, instrument parameter tables, analysis reports | All MC functions plus storage capability, expert query system with optimal condition settings, standard sample configuration, interference prompts, 8 working curves for comparison, standard point correction capability |
| Gas Control | Air acetylene flow pressure indication, secondary explosion-proof, multi-stage imported pressure stabilizing valve | Air acetylene flow pressure indication, secondary explosion-proof, multi-stage imported pressure stabilizing valve |